OAKTON PCD380 KIT

Supplied by Antylia (Formerly Cole Parmer)


  • 35660-80 | OAKTON PCD380 KIT
  • 35660-80 | OAKTON PCD380 KIT
Average lead time: Call for Lead Time
Need Assistance?

Jump to Section
Product Description
OAKTON PCD380 KIT
Specifications
Inventory Number:
COL3566080EA
Part Number:
35660-80
Supplier:
Antylia (Formerly Cole Parmer)
Supplier's Lead Time:
Call for Lead Time
Quantity:
1 per EA
Contract Availability:
Country of Origin:
Accessories Included:
pH sensor head, pH sensor cartridge, pH buffer set, EC sensor head, EC sensor cartridge, conductivity standards set, DO sensor with cap, four AA batteries, wireless SD card, and carrying case
Battery:
Four AA batteries
Buffer Recognition:
USA, DIN, NIST, NIST (10), and custom
Calibration:
pH: Up to 6 points; Conductivity: Up to 4 automatic, up to 5 manual
Conductivity Accuracy:
±0.5% full scale of each range, ±1.5% full scale at 200 mS
Conductivity Resolution:
Auto ranging, up to 4 significant digits
Connections:
micro USB, phono jack, and push-pull connection
Data Logging Points:
10000
Display Type:
Color graphic LCD with backlight
Dissolved Oxygen Accuracy:
±0.2 mg/L,±0.2%
Dissolved Oxygen Resolution:
0.01 mg/L, 0.1%
Height Cm:
22
Height In:
8.66
Interface:
USB and Wi-Fi
Max Conductivity:
200 S
Max Dissolved Oxygen:
20 mg/L, 200%
Max Mv:
1000
Max pH:
20
Max Resistivity:
200 M?•cm
Max Salinity:
80 ppt
Max Tds:
200,000 mg/L
Max Temperature C:
130
Max Temperature F:
266
Meter Type:
Portable
Min Conductivity:
0 S
Min Dissolved Oxygen:
0 mg/L, 0%
Min Mv:
-1000
Min Ph:
-2
Min Resistivity:
0.1 ?•cm
Min Salinity:
0 ppt
Min Tds:
0.01 mg/L
Min Temperature C:
-30
Min Temperature F:
-22
Mv Accuracy:
±0.1
Mv Resolution:
0.1
Nist Calibration Included:
No
Number Of Positions:
2
Output:
micro USB for PC connection, phono jack for printer connection
Ph Accuracy:
±0.005
Ph Resolution:
0.001
Probe Included:
Yes
Resistivity Accuracy:
±0.5% full scale of each range, ±1.5% full scale > 20.0M?
Resistivity Resolution:
Auto ranging, up to 4 significant digits
Salinity Accuracy:
±0.5% of reading or ±0.01 ppt, whichever is greater
Salinity Resolution:
0.01 ppt
Tds Accuracy:
±0.5% of reading or ±0.1 mg/L, whichever is greater
Tds Resolution:
0.01 minimum, 4 significant digits
Temperature Accuracy:
±0.9°F (±0.5°C)
Temperature Coefficient:
0.00 to 10.00%/°C
Temperature Compensation:
Automatic
Temperature Resolution:
0.1°F/°C
Waterproof:
IP67
Width cm:
3.45
Width in:
1.4
Additional Information